Last seen: May 16, 2024
If the Operating status shows error code of 10 or 11, the dropout issue may be fixed remotely. Please download the file below and follow the instructi...
The Mag Status table can be found on page 8 of the MFAM module user guide, which is inside the USB drive shipped with the MFAM. The user guide can b...
@kuldeep, No, the data streaming is ONLY through the Ethernet port, not the UART5 port. Current instrument software does not support the UART5 streami...
@ellipsis You are welcome! From the front of the front eyelets to the rear of the rear eyelets, it is a little more than 27 inches. So, your CAD readi...
@ellipsis We would recommend the rigid part to be ~ 1m. So with the landing gear, the MagArrow is about 1.5m from the drone.
Yes, it is UTC time.
@aldsantos I see. Please try it anyway next time in the field. Let us know if loss of connection still occurs after you have selected to stay connecte...
@aldsantos For the current version of the embedded software, some raw data was first stored in the microSD and then transferred to the tablet. Data lo...
Regarding your SD card question, you must mean the USB drive, right? The drive is not for storing mag data but for running the embedded software.
The reason why the tablet lost connection with the sensor is because the G-864 WIFI has no internet access. The tablet will try to switch to another W...
A CAD file of MagArrow can be downloaded here if you want to design such a mounting structure. Please note that the external helical antenna is NO...
A good suggestion! We will let you know how to do it soon.
For G-864, data from the sensor is transferred to the datalogger, which is a tablet, wirelessly via WIFI. As of now, users cannot transfer data from t...